ANDOR SHOMROCK SPECTROMETER & DATA
ACQUCITION SYSTEM
USER MANUAL
By Tao
Version 1
2011/10/21
目录
一. 探测器简介--------------------------------------------------------------------------------------------------3
1. 硅探测器(Dsi )--------------------------------------------------------------------------------------3 2. 光电倍增管(PMT )---------------------------------------------------------------------------------3 3. 铟镓砷探测器(DInGaAs )-------------------------------------------------------------------------3 4. 硫化铅探测器(DPbs )------------------------------------------------------------------------------5 5. 化铟探测器(DInSb)--------------------------------------------------------------------------------5 二. 单色仪简介--------------------------------------------------------------------------------------------------7 三. 数据采集装置-----------------------------------------------------------------------------------------------8
1.参数------------------------------------------------------------------------------------------------------8 2.前面板---------------------------------------------------------------------------------------------------9 3.后面板---------------------------------------------------------------------------------------------------9 四.整套系统实际使用---------------------------------------------------------------------------------------10
1.软件安装-----------------------------------------------------------------------------------------------10 2.连接-----------------------------------------------------------------------------------------------------10
(1)无SR830+SR540-------------------------------------------------------------------------10 (2)系统中配置了SR830+SR540----------------------------------------------------------11 3.软件使用-----------------------------------------------------------------------------------------------13 五.注意事项-------------------------------------------------------------------------------------------------19
一. 探测器简介 单通道光电探测器, 探测范围覆盖了紫外-可见光-红外(UV-Visible-IR )的宽波段范围。 1. 硅探测器(Dsi )
响应波段为200-1100nm,覆盖了UV-VIS-Near Infrared 波段,响应曲线在整个波段较平滑。不需要外接电源,直接接到光谱仪输出狭缝既可,灵敏度不高。
2. 光电倍增管(PMT )
响应波段为185-900nm(示不同型号而定),主要为UV-VIS 波段探测,输出信号为负极性(N), 有电流或电压型两种输出,具体示型号而定。需与高压电源搭配使用,高压电源电压可调节范围为0~1250V,电压越高探测器灵敏度越高,使用时首先根据图示选择为Local 模式,而后按Start HV按钮打开高压,在未知信号强度之前,建议从0V 慢慢通过旋转HV Adjustment旋钮慢慢升高电压,直到得到合适的信号强度,以保护探测器。
3. 铟镓砷探测器(DInGaAs )
响应波段为0.8~2600nm,示具体型号而定(可参照下表),目前Andor 提供DInGaAs 探测器均为TE 制冷型,工作时需要与所配制冷器共同使用已达到最佳效果,待制冷器上Stable 指示灯变为红色即代表探测器以达到指定温度,方可开始使用。
Parameters
Active diameter Wavelength range (nm) peak
responsivity (A/W) D*(typ.) NEP(typ.) Cooled Operating temp Temperature stability (℃) ambient temperature 信号输出模式 信号输出极性
InGaAs1700
InGaAs1900
InGaAs2200
InGaAs2400
InGaAs2600
800-1700
800-1900
800-2200
800-2400
800-2600
8.4×1013
9.1×1012
1.9×10121.4×10-13
ZTC
9.6×10112.8×10-13
ZTC
4.9×1011 5.5×10-13 ZTC
3.2×10-15-14
ZTC
ZTC
±0.5
±0.5
±0.5 +10~+40电流 (P )
±0.5 +10~+40电流 (P )
±0.5 电流 (P )
电流 (P )
电流 (P )
4. 硫化铅探测器(DPbs )
响应波段为0.8-3.2um(示不同型号而定), 覆盖范围为近红外到中红外波段
DPbs2900
DPbs3200
光敏面尺寸 mm 波长范围 μm 峰值波长 μm 响应Su V/W 电阻Rd MΩ D* cm(Hz)1/2/W 时间常数 μs 放大倍数
输入端失调电压 µV 前放输入端的漂移 µV频率响应范围 Hz 信号输出模式 输出信号极性
电压 正(P )
0.8
~2.9 ≥2.2 ≥3×104 ≥5×108 ≤200
0.8~3.2 ≥2.1 ≥300 ≥1×108 ≤400 <±1 ±1
100—1000(推荐400Hz )
电压 正(P )
×1,×10,×100
DPbS2900和DPbS3200硫化铅探测器为光导型红外探测器,使用时必须配合锁相放大器 通常使用由SRS 公司生产的 SR830锁相放大器.
5. 锑化铟探测器(DInSb)
DInSb5-DE(X) DInSb5-HS
◆ DInSb5-De(x)为液氮制冷型,x-01/ 02/ 04/ 07,四种光敏面尺寸可选,适合一般测量 ◆ DInSb5-HS为液氮制冷高速响应型,集成前置放大器,响应时间小于25ns
因此探测器噪声较高,工作时需要用液氮制冷,在其杜瓦罐被充满情况下可以工作8小时左右。最好与SRS 公司生产的SR830锁相放大器连用以提高灵敏度。在测试荧光寿命时可以连接示波器直接测量。
型号/参数 光敏面尺寸(mm) 波长范围(μm) 峰值响应度(A/W) 峰值响应度(V/W) 响应时间(ns) D*(@λpeak, 1KHz)cm Hz 1/2W -1 NEP(@λpeak, 1KHz)pW/Hz1/2暗电流(μA) 前置放大器 信号输出模式
0.8 7 选配 电流
1.6 30 选配 电流
3 110 选配 电流
6 350 选配 电流
- - 集成 电压
1 x 1011
1 x 1011
1 x 1011
1 x 1011
1 x 1011
-----
-
-
-
2x 104
Φ1Φ2DInSb5-De04
Φ4DInSb5-De07
Φ7DInSb5-HS 1×1(方)输出信号极性
正(P ) 正(P ) 正(P ) 正(P ) 正(P )
二. 单色仪简介
Figue 1 光谱仪外观图
单色仪内部结构图及简单功能原理
Figue 2 内部光路图
样品发射荧光通过入口狭缝进入光谱仪后打到准直镜上(从下数第一面黄色镜子),然后样品发光会到达光栅(图中蓝色三角形),经过光栅的复色光会被分光,不同波长的光会被反射到不同的方向上,经过聚焦透镜(从下数第二面镜子)聚焦到出口狭缝被探测器所探测。图中所示1,2为entrance port,1(side ),2(front )。A ,B 为exit port,A (front )B (side )
日常维护,因光谱仪内部多为光学器件,所以要避免使强光(例如激光)直接进入光谱仪,否则可能会损坏光谱仪内部光学元件。
三. 数据采集装置
Figue 3
此数据采集器是一个低噪声,高增益数据获取装置,用于与Andor 生产单点探测器匹配使用进行模数转换。
1.参数
CHA 、CHB Input Signal AUX Input Range AD590 A/D Converter Accuracy HVC1005 CHOPPER D/A Converter Accuracy
Current Voltage
±100nA ~±10mA ±10μV ~±10V
DC 0~10V -30~100 ℃
16 Bit DC 0~10V 12 Bit
Input ,2 Output
SOLENOID EX TRIG Data Points Communicating Interface
Apply to M30134-282R Solenoid
TTL rising edge ≤2000( Stored in RAM)
USB2.0
Power Supply
AC 100-240V ,50/60Hz(Y-GE-240-270)
2.前面板
(1) (2) (3) (4)
CHA :可接收电流或者电压信号 CHB :可接收电流或者电压信号 OVLD :通道A/B 过载指示灯
AUX :DC 0~10V输出,可通过软件控制输出电压实现对其他仪
器的控制
(5) (6) 3.后面板
AD590:interface POWER
(1) (2) (3) (4)
HVC :高压电源输出 Chopper::斩波器控制输出 Shutter::TTL 快门控制输出
Solenoid::螺线管输出, 可连接 M30134-282R
(5) (6) (7)
EX TRIG:外触发输入 DC Power:直流24v 电源输入
I/O Port:
5 input, 2 output, defined as table below:
四.整套系统实际使用 1.软件安装
此套系统使用软件为
Andor Solis Scanning
点击next ,然后选择安装路径即可完成安装。
在软件安装完成后方可把单色仪和数据采集装置连接到计算机UBS 接口,并接通电源,此时计算机会提示找到新硬件,一般自动安装即可。如不能安装,请联系服务工程师。 2.连接
(1)无SR830+SR540
Andor 单点探测器所配探测器均为BNC 输出,使用时通过BNC CABLE把探测器后方的BNC 接头与数据采集装置CHA 通道相连接。如有两个探测器则可以把另一个探测器连接到CHB 。
*如果探测器本身需要额外电源,则需要事先连接好。
(2)系统中配置了SR830+SR540
SR830为美国SRS 公司生产锁相放大器,与SR540连用可以有效提高整套系统灵敏度。
如果系统中有锁相放大系统,则需要把从探测器出来的信号连接到SR830输入端A/I,而后通过BNC CABLE 把前面板CH1 Output 信号输出到数据采集装置上。再用BNC 电缆把斩波器控制盒上的参考频率输出信号连接到锁相放大器(SR830)Ref In接口上,测量时一定需要使被探测光信号通过斩波器,此时锁相只接收与斩波器同一频率的信号,而自然光等其他噪声信号则被滤掉了。
斩波器的频率设置无固定要求,但最好不要选择50HZ 或者其倍数。 锁相放大器简单设置,根据探测器输出类型选择合适的锁相输入类型(V/I),
同时在图中标注部分选择R 和⊙ ,同时按output 按钮选择成Display 即可。因为被探测光频率由斩波器给出,所以需要选择外同步,具体请参照下图,按source
健即可切换。
测量时主要通过更改sensitivity 找到合适的信号值,注意不要使OVLD
灯亮。
3.软件使用
(1)确保打开软件之前所有仪器的电源被接通以及驱动程序已经被安
装正确.
(2)打开软件点击
Hardware→Select Hardware
根据所配置光谱仪数量及数据采集装置勾选恰当的设备
完成后点击OK
(3)点击
Hardware→Communication port
勾选上相应设备USB Mode 前的Enable,并选择好对应的序列号即可。
(4)点击Hardware→Connect Data Acquisition和 Connect Spectra A
如果正常连接,则可以在软件下方的 Information panel 处看到与仪器连接正常提示。
(5)实验设置
(ⅰ)点击
Options→Instrument→Data Capture Setting
之后会弹出以下窗口
Integration time(ms) 采样间隔,此时间为光谱仪每一个设置波长的采样时间
间隔,不宜过小,通常为100,完成后点击SET。如果与
SR830连用,则需要至少大于SR830 的 time constant
数值。
Trig Delaytime(us) 触发延时,通常不需要设置。
Channel No 选择采样的通道,根据需要使用的探测通道选择。 Gain A A通道增益,如果探测器为电流输出则选择范围为3-8,
电压范围为0-5.
IVINPUT A A通道信号输入类型,根据不同探测器输出类型选择电
流(I)或者电压(V)
POLARTINPUT A A通道极性, 根据不同探测器输出类型学则正(P)或
者负(N)
HVC(mv) 光电倍增管高压电源远程控制电压控制
SHUTTER ON 打开光谱仪快门(需配置)
SHUTTER OFF 关闭光谱仪快门(需配置)
Chopper(mv) 斩波器频率远程电压控制
Gain B, IVINPUT B, POLARTINPUT B 与A 通道相同
(ⅱ)在完成探测器极性以及I/V
输入选择完成后即可开始测试。
Ⅰ.实验时首先点击图中6位置选择好样品可能的发光波长位置以及需要选择的光栅(根据不同闪耀波长,如不确定可以点击OPTIONS→Instrument→Spectra A setting→Grating information查看 ),点Go,待光谱仪转动到相应位置后即可。
Ⅱ.确认探测器安装的出口探测器位置选择合适的出口,首先在第一个方框中选择需要控制的是Entrance/Exit Mirror,在第2个方框中选择Front/Side(可参考光谱仪简介)
Ⅲ.点击图中1位置的按钮进行按时间扫描,此时关注Signal Value 值,如果选择波长确实是样品发光,则应该可以看到数值,此时调节通过调节光谱仪入/出口狭缝得到一个比较理想的数值即可,此套仪器的满量程为65535,如果检测峰值波长位置数值可以达到50000左右最好。对于位置样品不熟悉发光波长,则可以小一些,避免出现非预期信号过大,损坏探测器。
(ⅲ)按波长扫描
在调节完信号大小后可以双击Operation parameter 选择测试光谱时
的相关参数。
首先点击Insert 之后输入相关参数,点击SAVE,而后点击SCAN 即可开始测试
Device Select 选择需要控制的光谱仪
Grating Select 根据闪耀波长及需要扫描光谱的精细程度选择合适的光栅,
光栅刻画线越密则可得到光谱越精细,闪耀波长则决定了光
栅在那个波段反射率越高。
Start(nm) 起始波长
Stop (nm) 截至波长
Step (nm) 扫描间隔
Average 每一个点的平均次数
Run Times 运行次数
Delay(us) 延迟时间,通常不需要选择
Detector type 软件内置探测器类型,根据探测器型号选择。如果与SR830
连用,则需要选择一个内置为V-P 模式的探测器。
Channel No. 需要探测的通道
Averaging Time 此时间为光谱仪每一个设置波长的采样时间间隔,不宜过小,
通常为100,完成后点击SET。如果与SR830连用,则需要至
少大于
SR830 的 Time constant 数值。
五.注意事项
(1)扫描光谱时一定注意不要使探测饱和,否则有可能损坏探测器,
如果由于信号过强导致探测被损坏,则不再保修范围。尤其注意大功率激光不要直入射到光谱仪,否则有可能会损坏光谱仪内部光学器件。
(2)扫描时注意激发光的二级谱线,因激发光强度远远大于样品发射
荧光,轻者会造成测得光谱很难看,严重则会损坏探测,需避免。通常选用高通滤光片放到光谱仪入口处过滤掉激发光。
(3)如果欲详细了解SR830及SR540请查看对应仪器的相关英文手册。
ANDOR SHOMROCK SPECTROMETER & DATA
ACQUCITION SYSTEM
USER MANUAL
By Tao
Version 1
2011/10/21
目录
一. 探测器简介--------------------------------------------------------------------------------------------------3
1. 硅探测器(Dsi )--------------------------------------------------------------------------------------3 2. 光电倍增管(PMT )---------------------------------------------------------------------------------3 3. 铟镓砷探测器(DInGaAs )-------------------------------------------------------------------------3 4. 硫化铅探测器(DPbs )------------------------------------------------------------------------------5 5. 化铟探测器(DInSb)--------------------------------------------------------------------------------5 二. 单色仪简介--------------------------------------------------------------------------------------------------7 三. 数据采集装置-----------------------------------------------------------------------------------------------8
1.参数------------------------------------------------------------------------------------------------------8 2.前面板---------------------------------------------------------------------------------------------------9 3.后面板---------------------------------------------------------------------------------------------------9 四.整套系统实际使用---------------------------------------------------------------------------------------10
1.软件安装-----------------------------------------------------------------------------------------------10 2.连接-----------------------------------------------------------------------------------------------------10
(1)无SR830+SR540-------------------------------------------------------------------------10 (2)系统中配置了SR830+SR540----------------------------------------------------------11 3.软件使用-----------------------------------------------------------------------------------------------13 五.注意事项-------------------------------------------------------------------------------------------------19
一. 探测器简介 单通道光电探测器, 探测范围覆盖了紫外-可见光-红外(UV-Visible-IR )的宽波段范围。 1. 硅探测器(Dsi )
响应波段为200-1100nm,覆盖了UV-VIS-Near Infrared 波段,响应曲线在整个波段较平滑。不需要外接电源,直接接到光谱仪输出狭缝既可,灵敏度不高。
2. 光电倍增管(PMT )
响应波段为185-900nm(示不同型号而定),主要为UV-VIS 波段探测,输出信号为负极性(N), 有电流或电压型两种输出,具体示型号而定。需与高压电源搭配使用,高压电源电压可调节范围为0~1250V,电压越高探测器灵敏度越高,使用时首先根据图示选择为Local 模式,而后按Start HV按钮打开高压,在未知信号强度之前,建议从0V 慢慢通过旋转HV Adjustment旋钮慢慢升高电压,直到得到合适的信号强度,以保护探测器。
3. 铟镓砷探测器(DInGaAs )
响应波段为0.8~2600nm,示具体型号而定(可参照下表),目前Andor 提供DInGaAs 探测器均为TE 制冷型,工作时需要与所配制冷器共同使用已达到最佳效果,待制冷器上Stable 指示灯变为红色即代表探测器以达到指定温度,方可开始使用。
Parameters
Active diameter Wavelength range (nm) peak
responsivity (A/W) D*(typ.) NEP(typ.) Cooled Operating temp Temperature stability (℃) ambient temperature 信号输出模式 信号输出极性
InGaAs1700
InGaAs1900
InGaAs2200
InGaAs2400
InGaAs2600
800-1700
800-1900
800-2200
800-2400
800-2600
8.4×1013
9.1×1012
1.9×10121.4×10-13
ZTC
9.6×10112.8×10-13
ZTC
4.9×1011 5.5×10-13 ZTC
3.2×10-15-14
ZTC
ZTC
±0.5
±0.5
±0.5 +10~+40电流 (P )
±0.5 +10~+40电流 (P )
±0.5 电流 (P )
电流 (P )
电流 (P )
4. 硫化铅探测器(DPbs )
响应波段为0.8-3.2um(示不同型号而定), 覆盖范围为近红外到中红外波段
DPbs2900
DPbs3200
光敏面尺寸 mm 波长范围 μm 峰值波长 μm 响应Su V/W 电阻Rd MΩ D* cm(Hz)1/2/W 时间常数 μs 放大倍数
输入端失调电压 µV 前放输入端的漂移 µV频率响应范围 Hz 信号输出模式 输出信号极性
电压 正(P )
0.8
~2.9 ≥2.2 ≥3×104 ≥5×108 ≤200
0.8~3.2 ≥2.1 ≥300 ≥1×108 ≤400 <±1 ±1
100—1000(推荐400Hz )
电压 正(P )
×1,×10,×100
DPbS2900和DPbS3200硫化铅探测器为光导型红外探测器,使用时必须配合锁相放大器 通常使用由SRS 公司生产的 SR830锁相放大器.
5. 锑化铟探测器(DInSb)
DInSb5-DE(X) DInSb5-HS
◆ DInSb5-De(x)为液氮制冷型,x-01/ 02/ 04/ 07,四种光敏面尺寸可选,适合一般测量 ◆ DInSb5-HS为液氮制冷高速响应型,集成前置放大器,响应时间小于25ns
因此探测器噪声较高,工作时需要用液氮制冷,在其杜瓦罐被充满情况下可以工作8小时左右。最好与SRS 公司生产的SR830锁相放大器连用以提高灵敏度。在测试荧光寿命时可以连接示波器直接测量。
型号/参数 光敏面尺寸(mm) 波长范围(μm) 峰值响应度(A/W) 峰值响应度(V/W) 响应时间(ns) D*(@λpeak, 1KHz)cm Hz 1/2W -1 NEP(@λpeak, 1KHz)pW/Hz1/2暗电流(μA) 前置放大器 信号输出模式
0.8 7 选配 电流
1.6 30 选配 电流
3 110 选配 电流
6 350 选配 电流
- - 集成 电压
1 x 1011
1 x 1011
1 x 1011
1 x 1011
1 x 1011
-----
-
-
-
2x 104
Φ1Φ2DInSb5-De04
Φ4DInSb5-De07
Φ7DInSb5-HS 1×1(方)输出信号极性
正(P ) 正(P ) 正(P ) 正(P ) 正(P )
二. 单色仪简介
Figue 1 光谱仪外观图
单色仪内部结构图及简单功能原理
Figue 2 内部光路图
样品发射荧光通过入口狭缝进入光谱仪后打到准直镜上(从下数第一面黄色镜子),然后样品发光会到达光栅(图中蓝色三角形),经过光栅的复色光会被分光,不同波长的光会被反射到不同的方向上,经过聚焦透镜(从下数第二面镜子)聚焦到出口狭缝被探测器所探测。图中所示1,2为entrance port,1(side ),2(front )。A ,B 为exit port,A (front )B (side )
日常维护,因光谱仪内部多为光学器件,所以要避免使强光(例如激光)直接进入光谱仪,否则可能会损坏光谱仪内部光学元件。
三. 数据采集装置
Figue 3
此数据采集器是一个低噪声,高增益数据获取装置,用于与Andor 生产单点探测器匹配使用进行模数转换。
1.参数
CHA 、CHB Input Signal AUX Input Range AD590 A/D Converter Accuracy HVC1005 CHOPPER D/A Converter Accuracy
Current Voltage
±100nA ~±10mA ±10μV ~±10V
DC 0~10V -30~100 ℃
16 Bit DC 0~10V 12 Bit
Input ,2 Output
SOLENOID EX TRIG Data Points Communicating Interface
Apply to M30134-282R Solenoid
TTL rising edge ≤2000( Stored in RAM)
USB2.0
Power Supply
AC 100-240V ,50/60Hz(Y-GE-240-270)
2.前面板
(1) (2) (3) (4)
CHA :可接收电流或者电压信号 CHB :可接收电流或者电压信号 OVLD :通道A/B 过载指示灯
AUX :DC 0~10V输出,可通过软件控制输出电压实现对其他仪
器的控制
(5) (6) 3.后面板
AD590:interface POWER
(1) (2) (3) (4)
HVC :高压电源输出 Chopper::斩波器控制输出 Shutter::TTL 快门控制输出
Solenoid::螺线管输出, 可连接 M30134-282R
(5) (6) (7)
EX TRIG:外触发输入 DC Power:直流24v 电源输入
I/O Port:
5 input, 2 output, defined as table below:
四.整套系统实际使用 1.软件安装
此套系统使用软件为
Andor Solis Scanning
点击next ,然后选择安装路径即可完成安装。
在软件安装完成后方可把单色仪和数据采集装置连接到计算机UBS 接口,并接通电源,此时计算机会提示找到新硬件,一般自动安装即可。如不能安装,请联系服务工程师。 2.连接
(1)无SR830+SR540
Andor 单点探测器所配探测器均为BNC 输出,使用时通过BNC CABLE把探测器后方的BNC 接头与数据采集装置CHA 通道相连接。如有两个探测器则可以把另一个探测器连接到CHB 。
*如果探测器本身需要额外电源,则需要事先连接好。
(2)系统中配置了SR830+SR540
SR830为美国SRS 公司生产锁相放大器,与SR540连用可以有效提高整套系统灵敏度。
如果系统中有锁相放大系统,则需要把从探测器出来的信号连接到SR830输入端A/I,而后通过BNC CABLE 把前面板CH1 Output 信号输出到数据采集装置上。再用BNC 电缆把斩波器控制盒上的参考频率输出信号连接到锁相放大器(SR830)Ref In接口上,测量时一定需要使被探测光信号通过斩波器,此时锁相只接收与斩波器同一频率的信号,而自然光等其他噪声信号则被滤掉了。
斩波器的频率设置无固定要求,但最好不要选择50HZ 或者其倍数。 锁相放大器简单设置,根据探测器输出类型选择合适的锁相输入类型(V/I),
同时在图中标注部分选择R 和⊙ ,同时按output 按钮选择成Display 即可。因为被探测光频率由斩波器给出,所以需要选择外同步,具体请参照下图,按source
健即可切换。
测量时主要通过更改sensitivity 找到合适的信号值,注意不要使OVLD
灯亮。
3.软件使用
(1)确保打开软件之前所有仪器的电源被接通以及驱动程序已经被安
装正确.
(2)打开软件点击
Hardware→Select Hardware
根据所配置光谱仪数量及数据采集装置勾选恰当的设备
完成后点击OK
(3)点击
Hardware→Communication port
勾选上相应设备USB Mode 前的Enable,并选择好对应的序列号即可。
(4)点击Hardware→Connect Data Acquisition和 Connect Spectra A
如果正常连接,则可以在软件下方的 Information panel 处看到与仪器连接正常提示。
(5)实验设置
(ⅰ)点击
Options→Instrument→Data Capture Setting
之后会弹出以下窗口
Integration time(ms) 采样间隔,此时间为光谱仪每一个设置波长的采样时间
间隔,不宜过小,通常为100,完成后点击SET。如果与
SR830连用,则需要至少大于SR830 的 time constant
数值。
Trig Delaytime(us) 触发延时,通常不需要设置。
Channel No 选择采样的通道,根据需要使用的探测通道选择。 Gain A A通道增益,如果探测器为电流输出则选择范围为3-8,
电压范围为0-5.
IVINPUT A A通道信号输入类型,根据不同探测器输出类型选择电
流(I)或者电压(V)
POLARTINPUT A A通道极性, 根据不同探测器输出类型学则正(P)或
者负(N)
HVC(mv) 光电倍增管高压电源远程控制电压控制
SHUTTER ON 打开光谱仪快门(需配置)
SHUTTER OFF 关闭光谱仪快门(需配置)
Chopper(mv) 斩波器频率远程电压控制
Gain B, IVINPUT B, POLARTINPUT B 与A 通道相同
(ⅱ)在完成探测器极性以及I/V
输入选择完成后即可开始测试。
Ⅰ.实验时首先点击图中6位置选择好样品可能的发光波长位置以及需要选择的光栅(根据不同闪耀波长,如不确定可以点击OPTIONS→Instrument→Spectra A setting→Grating information查看 ),点Go,待光谱仪转动到相应位置后即可。
Ⅱ.确认探测器安装的出口探测器位置选择合适的出口,首先在第一个方框中选择需要控制的是Entrance/Exit Mirror,在第2个方框中选择Front/Side(可参考光谱仪简介)
Ⅲ.点击图中1位置的按钮进行按时间扫描,此时关注Signal Value 值,如果选择波长确实是样品发光,则应该可以看到数值,此时调节通过调节光谱仪入/出口狭缝得到一个比较理想的数值即可,此套仪器的满量程为65535,如果检测峰值波长位置数值可以达到50000左右最好。对于位置样品不熟悉发光波长,则可以小一些,避免出现非预期信号过大,损坏探测器。
(ⅲ)按波长扫描
在调节完信号大小后可以双击Operation parameter 选择测试光谱时
的相关参数。
首先点击Insert 之后输入相关参数,点击SAVE,而后点击SCAN 即可开始测试
Device Select 选择需要控制的光谱仪
Grating Select 根据闪耀波长及需要扫描光谱的精细程度选择合适的光栅,
光栅刻画线越密则可得到光谱越精细,闪耀波长则决定了光
栅在那个波段反射率越高。
Start(nm) 起始波长
Stop (nm) 截至波长
Step (nm) 扫描间隔
Average 每一个点的平均次数
Run Times 运行次数
Delay(us) 延迟时间,通常不需要选择
Detector type 软件内置探测器类型,根据探测器型号选择。如果与SR830
连用,则需要选择一个内置为V-P 模式的探测器。
Channel No. 需要探测的通道
Averaging Time 此时间为光谱仪每一个设置波长的采样时间间隔,不宜过小,
通常为100,完成后点击SET。如果与SR830连用,则需要至
少大于
SR830 的 Time constant 数值。
五.注意事项
(1)扫描光谱时一定注意不要使探测饱和,否则有可能损坏探测器,
如果由于信号过强导致探测被损坏,则不再保修范围。尤其注意大功率激光不要直入射到光谱仪,否则有可能会损坏光谱仪内部光学器件。
(2)扫描时注意激发光的二级谱线,因激发光强度远远大于样品发射
荧光,轻者会造成测得光谱很难看,严重则会损坏探测,需避免。通常选用高通滤光片放到光谱仪入口处过滤掉激发光。
(3)如果欲详细了解SR830及SR540请查看对应仪器的相关英文手册。